Heated Nitrogen Purifiers from VICI Valco Instruments

vici valco helium purifier

Carrier gas purity is essential in any application requiring extreme sensitivity. Impurities limit detector sensitivity and can even destroy capillary columns. The Valco nitrogen purifier provides "point-of-use" purification to sub-ppm levels of reactive gaseous impurities.

Based on 10 ppm total inlet impurities, outlet impurities are less than 10 ppb for H2O, H2, O2, NO, NH3, CO, and CO2. Other impurities removed include CF4, CCl4, SiH4, and light hydrocarbons. CH4, He, Ne, Ar, Kr, Xe, and Rn are not removed.

The purification substrate is a non-evaporable heat-activated gettering alloy. This stable alloy is contained in a welded assembly, so the purifiers can be used safely in industrial applications with minimal precautions. When the getter is heated, the oxide film on the particle surface is eliminated, allowing nitrogen to diffuse into the bulk of the getter particles. The purifier features a self-regulating design which maintains the getter material at the optimum temperature and eliminates the possibility of thermal runaway.

vici valco miniature helium purifier

The miniature version is designed to be installed in a gas chromatograph's flow path immediately upstream of the injector. The NPM will remove any contaminants introduced by flow controllers, elastomeric tube seals, pressure regulators, crude traps, or other system components that are not completely clean and leak-tight.

Purifers include a universal power supply.
Description Voltage Product No.
HEATED NITROGEN PURIFIERS
Standard size 110 VAC NP2
230 VAC NP2-220
Miniature 110 VAC NPM
230 VAC NPM-220
REPLACEMENT POWER SUPPLY
For standard and miniature purifiers 110 VAC PS24VDC-CE
230 VAC PS24VDC-CE-220
REPLACEMENT GETTER ASSEMBLY
For standard purifier only I-23572NP2

MORE INFORMATION

MAX OPERATING PRESSURE

Standard: 1000 psig

Minature: 200 psig

CE COMPLIANCE

DISPOSING OF SPENT GETTER CARTRIDGES

Contact VICI to obtain a return authorization number. The packaged getter cartridge should be clearly marked "Traps for Disposal".